Por favor, use este identificador para citar o enlazar este ítem:
https://hdl.handle.net/20.500.12008/30116
Cómo citar
Título: | High precision voltage ratio measurements using a 3458A multimeter |
Autor: | Slomovitz, Daniel Trigo, Leonardo |
Tipo: | Ponencia |
Palabras clave: | Calibration, Divider, Voltmeter, Uncertainty |
Fecha de publicación: | 2006 |
Resumen: | This paper shows that very high precision ratio measurements (uncertainty of 3 parts in 10⁶) can be performed, also at frequencies of 1 kHz, using the multimeter HP 3458A. It is proposed to slightly change known algorithms to control the instrument, and a new uncertainty calculation is presented. |
EN: | Conference on Precision Electromagnetic Measurement, CPEM Digest, Torino, Italy, jul. 2006, pp. 412-413. |
Citación: | Slomovitz, D. y Trigo, L. High precision voltage ratio measurements using a 3458A multimeter [en línea]. EN: Conference on Precision Electromagnetic Measurement, CPEM Digest, Torino, Italy, jul. 2006, pp. 412-413. |
Aparece en las colecciones: | Publicaciones académicas y científicas - Instituto de Ingeniería Eléctrica |
Ficheros en este ítem:
Fichero | Descripción | Tamaño | Formato | ||
---|---|---|---|---|---|
ST06.pdf | Versión definitiva | 1,62 MB | Adobe PDF | Visualizar/Abrir |
Este ítem está sujeto a una licencia Creative Commons Licencia Creative Commons