english Icono del idioma   español Icono del idioma  

Por favor, use este identificador para citar o enlazar este ítem: https://hdl.handle.net/20.500.12008/30116 Cómo citar
Título: High precision voltage ratio measurements using a 3458A multimeter
Autor: Slomovitz, Daniel
Trigo, Leonardo
Tipo: Ponencia
Palabras clave: Calibration, Divider, Voltmeter, Uncertainty
Fecha de publicación: 2006
Resumen: This paper shows that very high precision ratio measurements (uncertainty of 3 parts in 10⁶) can be performed, also at frequencies of 1 kHz, using the multimeter HP 3458A. It is proposed to slightly change known algorithms to control the instrument, and a new uncertainty calculation is presented.
EN: Conference on Precision Electromagnetic Measurement, CPEM Digest, Torino, Italy, jul. 2006, pp. 412-413.
Citación: Slomovitz, D. y Trigo, L. High precision voltage ratio measurements using a 3458A multimeter [en línea]. EN: Conference on Precision Electromagnetic Measurement, CPEM Digest, Torino, Italy, jul. 2006, pp. 412-413.
Licencia: Licencia Creative Commons Atribución - No Comercial - Sin Derivadas (CC - By-NC-ND 4.0)
Aparece en las colecciones: Publicaciones académicas y científicas - Instituto de Ingeniería Eléctrica

Ficheros en este ítem:
Fichero Descripción Tamaño Formato   
ST06.pdfVersión definitiva1,62 MBAdobe PDFVisualizar/Abrir


Este ítem está sujeto a una licencia Creative Commons Licencia Creative Commons Creative Commons