Por favor, use este identificador para citar o enlazar este ítem:
https://hdl.handle.net/20.500.12008/43523
Cómo citar
Título: | Total knee replacement evaluation with Dynamic CINARTRO Method |
Autor: | Artigas, José Santos, Alejandro Simini, Franco |
Tipo: | Artículo |
Palabras clave: | Functional dynamic imaging, X rays, Videofluoroscopy, Prosthesis, Total knee replacement |
Descriptores: | Sistemas y Control |
Fecha de publicación: | 2018 |
Resumen: | Videofluoroscopy is used to evaluate the movement of the knee. The use of CINARTRO is here extended to Total Knee Replacement follow up, after incipient use in Anterior Cruciate Ligament repair. Tibial Femoral Contact Point (TFCP)
migration is smaller (42%) in prosthetic knees with respect to normal knees (55%). Considering flexion-extension stages
separately, the TFCP migration at full extension has a greater value (64%) than at mid extension and flexion (50% & 54%). The same behaviour is also observed for prosthetic knees, but all values reduced: a high of 50% at extension follows 37% & 41%. This can be interpreted as a prosthetic restored near-physiological behaviour. |
Editorial: | Sociedad Argentina de Bioingeniería |
EN: | Revista Argentina de Bioingeniería, v 22, no, 1, 2018 |
e-ISSN: | 2591-376X |
Citación: | Artigas, J, Santos, A, Simini, F. "Total knee replacement evaluation with Dynamic CINARTRO Method" Revista Argentina de Bioingeniería, v 22, no, 1, 2018 |
Licencia: | Licencia Creative Commons Atribución - No Comercial - Sin Derivadas (CC - By-NC-ND 4.0) |
Aparece en las colecciones: | Publicaciones académicas y científicas - Instituto de Ingeniería Eléctrica |
Ficheros en este ítem:
Fichero | Descripción | Tamaño | Formato | ||
---|---|---|---|---|---|
ASS17.pdf | 311,9 kB | Adobe PDF | Visualizar/Abrir |
Este ítem está sujeto a una licencia Creative Commons Licencia Creative Commons