english Icono del idioma   español Icono del idioma  

Por favor, use este identificador para citar o enlazar este ítem: https://hdl.handle.net/20.500.12008/41754 Cómo citar
Registro completo de metadatos
Campo DC Valor Lengua/Idioma
dc.contributor.authorFichtinger, Gabores
dc.contributor.authorGómez, Alvaroes
dc.contributor.authorLasso, Andrases
dc.contributor.authorCarbajal, Guillermoes
dc.date.accessioned2023-12-11T19:57:36Z-
dc.date.available2023-12-11T19:57:36Z-
dc.date.issued2013es
dc.date.submitted20231211es
dc.identifier.citationCarbajal, G, Lasso, A, Gómez, Á. Fichtinger, G. "Improving N-wire phantom-based freehand ultrasound calibration". International Journal of Computer Assisted Radiology and Surgery 2013, no.8, pp. 1063–1072. https://doi.org/10.1007/s11548-013-0904-9es
dc.identifier.urihttps://hdl.handle.net/20.500.12008/41754-
dc.description.abstractPurpose Freehand tracked ultrasound imaging is an inexpensive non-invasive technique used in several guided interventions. This technique requires spatial calibration between the tracker and the ultrasound image plane. Several calibration devices (a.k.a. phantoms) use N-wires that are convenient for automatic procedures since the segmentation of fiducials in the images and the localization of the middle wires in space are straightforward and can be performed in real time. The procedures reported in literature consider only the spatial position of the middle wire. We investigate if better results can be achieved if the information of all the wires is equally taken into account. We also evaluated the precision and accuracy of the implemented methods to allow comparison with other methods. Methods We consider a cost function based on the in-plane errors between the intersection of all the wires with the image plane and their respective segmented points in the image. This cost function is minimized iteratively starting from a seed computed with a closed-form solution based on the middle wires. Results Mean calibration precision achieved with the N-wire phantom was about 0.5mm using a shallow probe, and mean accuracy was around 1.4mm with all implemented methods. Precision was about 2.0mm using a deep probe. Conclusions Precision and accuracy achieved with the N-wire phantom and a shallow probe are at least comparable to that obtained with other methods traditionally considered more precise. Calibration using N-wires can be done more consistently if the parameters are optimized with the proposed cost function.es
dc.languageenes
dc.rightsLas obras depositadas en el Repositorio se rigen por la Ordenanza de los Derechos de la Propiedad Intelectual de la Universidad De La República. (Res. Nº 91 de C.D.C. de 8/III/1994 – D.O. 7/IV/1994) y por la Ordenanza del Repositorio Abierto de la Universidad de la República (Res. Nº 16 de C.D.C. de 07/10/2014)es
dc.subjectFreehand ultrasoundes
dc.subjectCalibrationes
dc.subjectN-wire phantomes
dc.subjectAccuracyes
dc.subject.otherProcesamiento de Señaleses
dc.titleImproving N-Wire Phantom-based Freehand Ultrasound Calibrationes
dc.typePreprintes
dc.rights.licenceLicencia Creative Commons Atribución - No Comercial - Sin Derivadas (CC - By-NC-ND 4.0)es
Aparece en las colecciones: Publicaciones académicas y científicas - Instituto de Ingeniería Eléctrica

Ficheros en este ítem:
Fichero Descripción Tamaño Formato   
CLGF13b.pdf868,4 kBAdobe PDFVisualizar/Abrir


Este ítem está sujeto a una licencia Creative Commons Licencia Creative Commons Creative Commons