english Icono del idioma   español Icono del idioma  

Por favor, use este identificador para citar o enlazar este ítem: https://hdl.handle.net/20.500.12008/41276 Cómo citar
Registro completo de metadatos
Campo DC Valor Lengua/Idioma
dc.contributor.authorPérez-Díaz, Leticia-
dc.contributor.authorGehrke, Sergio Alexandre-
dc.contributor.authorOdella Colla, Germán-
dc.contributor.authorDe Aza, Piedad N.-
dc.contributor.authorDedavid, Berenice Anina-
dc.date.accessioned2023-11-16T12:11:03Z-
dc.date.available2023-11-16T12:11:03Z-
dc.date.issued2022-
dc.identifier.citationPérez-Díaz, L, Gehrke, S, Odella Colla, G, [y otros autores]. "Titanium surface analysis after instrumentation with different burs simulating the implantoplasty technique: a pilot In vitro experimental study". Applied Science. [en línea] 2022, 12(15): 7920. 10 h. DOI: 10.3390/app12157920es
dc.identifier.issn2076-3417-
dc.identifier.urihttps://hdl.handle.net/20.500.12008/41276-
dc.description.abstractBackground: The present pilot in vitro study evaluated, physically and biologically, the effects produced by the wear of the titanium surface using different drill models. (2) Methods: Titanium disks were subjected to wear using four different burs and accordingly divided into the following test groups (n = 12 disks per group): Tungsten Burs (TB group), Tungsten Carbide Burs (TCB group), Coarse-Grained Diamond Burs (CGB group), and Fine-Grained Diamond Burs (FGB group). As a control group (CON group), titanium disks with a smooth surface (machined) were used. The samples were subjected to atomic force microscopy (AFM), profilometry analysis, scanning electron microscopy (SEM), and cell viability and adhesion assessments. (3) Results: The values of the measured roughness parameters showed statistical differences among the five groups (p = 0.0035 for Ra and p = 0.0010 for Rz). All test groups showed an important difference statistically (p = 0.0032) to the CON group for the cell viability and adhesion analysis. The data of cell absorbance at 570 nm were 0.4122 ± 0.05 for the CON group, 0.1354 ± 0.02 for the TB group, 0.123 ± 0.01 for the TCB group, 0.1414 ± 0.02 for the CGB group, and 0.1216 ± 0.03 for the FGB group. Additionally, the cell count showed the following adherence percentages: 57.6 ± 4.6% for the CON group, 22.9 ± 3.3% for the TB group, 23.4 ± 2.9% for the TCB group, 22.5 ± 3.1% for the CGB group, and 23.7 ± 3.3% for the FGB group. However, no statistical differences were found among the four test groups analyzed (p = 0.3916). (4) Conclusions: The results showed that the changes produced on the surface by the four different bur models altered the topography characteristics and affected the cell viability and adhesion in comparison with the control group.es
dc.format.extent10 h.es
dc.format.mimetypeapplication/pdfes
dc.language.isoen_USes
dc.publisherMDPIes
dc.relation.ispartofApplied Science, 2022, 12(15): 7920.es
dc.rightsLas obras depositadas en el Repositorio se rigen por la Ordenanza de los Derechos de la Propiedad Intelectual de la Universidad de la República.(Res. Nº 91 de C.D.C. de 8/III/1994 – D.O. 7/IV/1994) y por la Ordenanza del Repositorio Abierto de la Universidad de la República (Res. Nº 16 de C.D.C. de 07/10/2014)es
dc.subjectTitanium surfacees
dc.subjectImplantoplastyes
dc.subjectCell viability and adhesiones
dc.subjectPeri-implantitises
dc.subjectTitanium instrumentationes
dc.subjectFibroblast cellses
dc.titleTitanium surface analysis after instrumentation with different burs simulating the implantoplasty technique: a pilot In vitro experimental studyes
dc.typeArtículoes
dc.contributor.filiacionPérez-Díaz Leticia, Universidad de la República (Uruguay). Facultad de Ciencias. Instituto de Química Biológica.-
dc.contributor.filiacionGehrke Sergio Alexandre-
dc.contributor.filiacionOdella Colla Germán-
dc.contributor.filiacionDe Aza Piedad N.-
dc.contributor.filiacionDedavid Berenice Anina-
dc.rights.licenceLicencia Creative Commons Atribución (CC - By 4.0)es
dc.identifier.doi10.3390/app12157920-
Aparece en las colecciones: Publicaciones académicas y científicas - Facultad de Ciencias

Ficheros en este ítem:
Fichero Descripción Tamaño Formato   
10.390app12157920.pdf2,71 MBAdobe PDFVisualizar/Abrir


Este ítem está sujeto a una licencia Creative Commons Licencia Creative Commons Creative Commons