dc.contributor.author | David, Marcelo | es |
dc.contributor.author | Pracca, Francisco | es |
dc.contributor.author | Simini, Franco | es |
dc.date.accessioned | 2023-11-14T17:04:15Z | - |
dc.date.available | 2023-11-14T17:04:15Z | - |
dc.date.issued | 2011 | es |
dc.date.submitted | 20231114 | es |
dc.identifier.citation | David, M., Pracca, F., Simini, F. "Non-invasive Negative Pressure System to Treat Abdominal Hypertension". Publicado en: Jobbágy, Á. (eds) 5th European Conference of the International Federation for Medical and Biological Engineering. IFMBE Proceedings, vol 37, 2011. Springer, Berlin, Heidelberg. https://doi.org/10.1007/978-3-642-23508-5_56 | es |
dc.identifier.uri | https://hdl.handle.net/20.500.12008/41095 | - |
dc.description | Trabajo presentado al 5th European Conference of the International Federation for Medical and Biological Engineering, Budapest, Hungría, 14-18 Sep. 2011. | es |
dc.description.abstract | Abdominal hypertension can complicate critical care conditions. If untreated and sustained, abdominal hypertension can lead to the abdominal compartment syndrome with its high morbidity and mortality. Treatment of+ severe abdominal hypertension is often surgical. We designed ABDOPRE, a system to non-invasively lower abdominal pressure. A bell-shaped chamber is applied on the patient’s abdomen and by means of a vacuum pump a negative pressure is generated in the chamber outside the abdomen. ABDOPRE senses intra-abdominal pressure from an indwelling urinary bladder catheter. The desired value of abdominal pressure and the time allowed to reach this value are entered into the system which servo-controls the vacuum pump, according to treatment protocols. Pressure data of the monitoring/treatment sessions are recorded to allow review of patient response and a daily report is available for the Electronic Medical Record. Preliminary observations were made on intensive care patients. | es |
dc.language | en | es |
dc.publisher | Udelar.FI | es |
dc.rights | Las obras depositadas en el Repositorio se rigen por la Ordenanza de los Derechos de la Propiedad Intelectual de la Universidad De La República. (Res. Nº 91 de C.D.C. de 8/III/1994 – D.O. 7/IV/1994) y por la Ordenanza del Repositorio Abierto de la Universidad de la República (Res. Nº 16 de C.D.C. de 07/10/2014) | es |
dc.title | Non-invasive negative pressure system to treat abdominal hypertension | es |
dc.type | Ponencia | es |
dc.rights.licence | Licencia Creative Commons Atribución - No Comercial - Sin Derivadas (CC - By-NC-ND 4.0) | es |