english Icono del idioma   español Icono del idioma  

Por favor, use este identificador para citar o enlazar este ítem: https://hdl.handle.net/20.500.12008/32480 Cómo citar
Título: Countertorque testing and histomorphometric analysis of various implant surfaces in canines: a pilot study.
Autor: Pebé, Pablo
Barbot, Roberto
Trinidad, Javier
Pesquera, A.
Lucente, J.
Nishimura, R.
Nasr, H.
Tipo: Artículo
Descriptores: IMPLANTACION DENTAL ENDOOSEA, GRABADO ACIDO DENTAL, ANALISIS DE VARIANZA, ANIMALES, MATERIALES DENTALES, PERROS, FEMUR, MICROSCOPIA ELECTRONICA DE RASTREO, OSEOINTEGRACION, TITANIO, PROPIEDADES DE SUPERFICIE
Fecha de publicación: 1997
Resumen: As surface roughness may play a role in the mechanical attachment of an implant surface to bone, various implant surfaces have been prepared and analyzed by removal torque (countertorque) or push-out tests in a variety of animal model systems. Rougher surfaces generally have displayed higher mechanical testing values, indicating a stronger implant-bone interface. This pilot study was undertaken to test the countertorque values for integrated threaded implants with surfaces prepared by machining, blasting, and acid-etching, to compare the various implant surface types histomorphometrically for percentage of bone-implant contact under loaded and unloaded conditions, and to determine the degree of correlation between countertorque values and bone-implant contact with varying degrees of surface roughness. The results of this animal investigation suggest that the strength of the bone-implant interface, as determined by countertorque testing, is influenced by different surface characteristics. Acid-etched surfaces resisted countertorque forces more successfully as compared with blasted or machined surfaces. Histologic evaluation of bone contact with the various implant surfaces did not demonstrate a definite advantage for rougher surfaces in regard to percentage of bone contact at the light microscopic level.
EN: Implant Dent ; 6(4): 259-65, 1997.
DOI: DOI: 10.1097/00008505-199700640-00002
Citación: Pebé, P, Barbot, R, Trinidad, J y otros."Countertorque testing and histomorphometric analysis of various implant surfaces in canines: a pilot study". Implant Dent ; 6(4): 259-65. [en línea] 1997 DOI: 10.1097/00008505-199700640-00002
Licencia: Licencia Creative Commons Atribución - No Comercial (CC - By-NC 4.0)
Aparece en las colecciones: Artículos y capítulos de libros hasta 2019 - Facultad de Odontología



Este ítem está sujeto a una licencia Creative Commons Licencia Creative Commons Creative Commons