english Icono del idioma   español Icono del idioma  

Por favor, use este identificador para citar o enlazar este ítem: https://hdl.handle.net/20.500.12008/21264 Cómo citar
Título: A fully integrated physical activity sensing circuit for implantable pacemakers
Autor: Arnaud, Alfredo
Galup Montoro, Carlos
Tipo: Artículo
Palabras clave: Analog design, CMOS, Low-power, Biomedical
Descriptores: ELECTRÓNICA
Fecha de publicación: 2004
Resumen: This paper shows the implementation of a fully integrated Gm-C 0.5-7Hz bandpass filter-amplifier with gain G=400, for a piezoelectric accelerometer which is part of a rate adaptive pacemaker. The fabricated circuit operates up to 2V power supply, consumes only 230nA current, and achives 2.1μVrms input noise. Detailed circuit specifications, measurements, and a comparative analysis of the system performance are presented.
Descripción: Postprint
Editorial: IEEE
Citación: Arnaud, A., Galup Montoro, C. A fully integrated physical activity sensing circuit for implantable pacemakers [en línea] SBCCI 2004. 17th Symposium on Integrated Circuits and Systems Design, Porto de Galinhas, Brasil, 2004. doi: 10.1109/SBCCI.2004.240770
Licencia: Licencia Creative Commons Atribución – No Comercial – Sin Derivadas (CC - By-NC-ND)
Aparece en las colecciones: Publicaciones académicas y científicas - Instituto de Ingeniería Eléctrica

Ficheros en este ítem:
Fichero Descripción Tamaño Formato   
AG04a.pdf375,94 kBAdobe PDFVisualizar/Abrir


Este ítem está sujeto a una licencia Creative Commons Licencia Creative Commons Creative Commons