Por favor, use este identificador para citar o enlazar este ítem:
https://hdl.handle.net/20.500.12008/21264
Cómo citar
| Título: | A fully integrated physical activity sensing circuit for implantable pacemakers |
| Autor: | Arnaud, Alfredo Galup Montoro, Carlos |
| Tipo: | Artículo |
| Palabras clave: | Analog design, CMOS, Low-power, Biomedical |
| Descriptores: | ELECTRÓNICA |
| Fecha de publicación: | 2004 |
| Resumen: | This paper shows the implementation of a fully integrated Gm-C 0.5-7Hz bandpass filter-amplifier with gain G=400, for a piezoelectric accelerometer which is part of a rate adaptive pacemaker. The fabricated circuit operates up to 2V power supply, consumes only 230nA current, and achives 2.1μVrms input noise. Detailed circuit specifications, measurements, and a comparative analysis of the system performance are presented. |
| Descripción: | Postprint |
| Editorial: | IEEE |
| Citación: | Arnaud, A., Galup Montoro, C. A fully integrated physical activity sensing circuit for implantable pacemakers [en línea] SBCCI 2004. 17th Symposium on Integrated Circuits and Systems Design, Porto de Galinhas, Brasil, 2004. doi: 10.1109/SBCCI.2004.240770 |
| Departamento académico: | Electrónica |
| Grupo de investigación: | Microelectrónica |
| Licencia: | Licencia Creative Commons Atribución – No Comercial – Sin Derivadas (CC - By-NC-ND) |
| Aparece en las colecciones: | Publicaciones académicas y científicas - Instituto de Ingeniería Eléctrica |
Ficheros en este ítem:
| Fichero | Descripción | Tamaño | Formato | ||
|---|---|---|---|---|---|
| AG04a.pdf | 375,94 kB | Adobe PDF | Visualizar/Abrir |
Este ítem está sujeto a una licencia Creative Commons Licencia Creative Commons