Por favor, use este identificador para citar o enlazar este ítem: https://hdl.handle.net/20.500.12008/37218

Título:

Trends in volumetric-energy efficiency of implantable neurostimulators : A review from a circuits and systems perspective.

Otros títulos:

Coordinador:

Director:

Compilador:

Autor:

Martínez Bentancor, Santiago
Veirano Núñez, Francisco
Constandinou, Timothy G.
Silveira, Fernando

Tutor:

Tipo de documento:

Preprint

Editor:

Palabras clave:

Batteries
Pain
Diseases
Medical treatment
Sleep apnea
Medical devices
Sensors
Neurostimulators
Neuromodulation
Implantable medical devices
Neural interface
Energy efficiency
Miniaturisation
Batt

Descriptores:

Año de publicación:

2023

Contenido:

Resumen:

This paper presents a comprehensive review of state-of-the-art, commercially available neurostimulators. We analyse key design parameters and performance metrics of 45 implantable medical devices across six neural target categories: deep brain, vagus nerve, spinal cord, phrenic nerve, sacral nerve and hypoglossal nerve. We then benchmark these alongside modern cardiac pacemaker devices that represent a more established market. This work studies trends in device size, electrode number, battery technology (i.e., primary and secondary use and chemistry), power consumption and longevity. This information is analysed to show the course of design decisions adopted by industry and identifying opportunity for further innovation. We identify fundamental limits in power consumption, longevity and size as well as the interdependencies and trade-offs. We propose a figure of merit to quantify volumetric efficiency within specific therapeutic targets, battery technologies/capacities, charging capabilities and electrode count. Finally, we compare commercially available implantable medical devices with recently developed systems in the research community. We envisage this analysis to aid circuit and system designers in system optimisation and identifying innovation opportunities, particularly those related to low power circuit design techniques.

Descripción:

Publicado en IEEE Transactions on Biomedical Circuits and Systems, vol. 17, no 1, feb 2023, pp 2-20

Editorial:

EN:

Financiadores:

CSIC - Universidad de la República.

Citación:

Martínez Bentancor, S., Veirano Núñez, F., Constandinou, T. y otros. "Trends in volumetric-energy efficiency of implantable neurostimulators : A review from a circuits and systems perspective" [Preprint]. Publicado en: IEEE Transactions on Biomedical Circuits and Systems. 2023, vol. 17, no 1, pp 2-20. DOI: 10.1109/TBCAS.2022.3228895

Citación:

metadata.cc.license.name:

Documento del Gobiberno:

URI:

https://ieeexplore.ieee.org/document/9984934
https://hdl.handle.net/20.500.12008/37218

ISBN:

e-ISBN:

ISSN:

1932-4545

ISMN:

Otros identificadores:

Título Obtenido:

Facultad o Servicio que otorga el Título:

Licencia:

Licencia Creative Commons Atribución - No Comercial - Sin Derivadas (CC - By-NC-ND 4.0)

Cobertura geográfica:

Cobertura temporal:

metadata.dc.relation.IsVersionOf:

Clasificación:

Es versión de:

FicherosDescripciónTamañoFormato
MVCS23.pdf Preprint 11.96 MB Adobe PDF