Título:
High precision voltage ratio measurements using a 3458A multimeter
Otros títulos:
Coordinador:
Director:
Compilador:
Autor:
Slomovitz, Daniel
Trigo, Leonardo
Trigo, Leonardo
Tutor:
Tipo de documento:
Ponencia
Editor:
Palabras clave:
Calibration
Divider
Voltmeter
Uncertainty
Divider
Voltmeter
Uncertainty
Descriptores:
Año de publicación:
2006
Contenido:
Resumen:
This paper shows that very high precision ratio measurements (uncertainty of 3 parts in 10⁶) can be performed, also at frequencies of 1 kHz, using the multimeter HP 3458A. It is proposed to slightly change known algorithms to control the instrument, and a new uncertainty calculation is presented.
Descripción:
metadata.articulos.dc.description.uri:
Editorial:
EN:
Conference on Precision Electromagnetic Measurement, CPEM Digest, Torino, Italy, jul. 2006, pp. 412-413.
Financiadores:
Citación:
Slomovitz, D. y Trigo, L. High precision voltage ratio measurements using a 3458A multimeter [en línea]. EN: Conference on Precision Electromagnetic Measurement, CPEM Digest, Torino, Italy, jul. 2006, pp. 412-413.