Por favor, use este identificador para citar o enlazar este ítem: https://hdl.handle.net/20.500.12008/30116

Título:

High precision voltage ratio measurements using a 3458A multimeter

Otros títulos:

Coordinador:

Director:

Compilador:

Autor:

Slomovitz, Daniel
Trigo, Leonardo

Tutor:

Tipo de documento:

Ponencia

Editor:

Palabras clave:

Calibration
Divider
Voltmeter
Uncertainty

Descriptores:

Año de publicación:

2006

Contenido:

Resumen:

This paper shows that very high precision ratio measurements (uncertainty of 3 parts in 10⁶) can be performed, also at frequencies of 1 kHz, using the multimeter HP 3458A. It is proposed to slightly change known algorithms to control the instrument, and a new uncertainty calculation is presented.

Descripción:

metadata.articulos.dc.description.uri:

Editorial:

EN:

Conference on Precision Electromagnetic Measurement, CPEM Digest, Torino, Italy, jul. 2006, pp. 412-413.

Financiadores:

Citación:

Slomovitz, D. y Trigo, L. High precision voltage ratio measurements using a 3458A multimeter [en línea]. EN: Conference on Precision Electromagnetic Measurement, CPEM Digest, Torino, Italy, jul. 2006, pp. 412-413.

Citación:

metadata.articulos.cc.license.name:

ISBN:

e-ISBN:

ISSN:

ISMN:

Otros identificadores:

Cobertura geográfica:

Cobertura temporal:

Departamento académico:

Grupo de investigación:

Licencia:

Licencia Creative Commons Atribución - No Comercial - Sin Derivadas (CC - By-NC-ND 4.0)
FicherosDescripciónTamañoFormato
ST06.pdf Versión definitiva 1.58 MB Adobe PDF